Pintuition Detector
2
Display
3
Volume Knob
4
Calibrate Button
5
Connector Port
10
Distance Measure
11
Pintuition Probe Status
12
Volume Status
13
Pintuition Probe
14
Pintuition Probe Cable
16
Probe Connector
16
14
2
Calibration Required
Calibration In Progress
Pintuition Probe not connected
Pintuition Probe connected
Audio muted
volume generates signal
outside of detec-
generates no
signal
Probe
SIRIUS MEDICAL SYSTEMS B.V.
|
High Tech Campus 41
|
5656AE
Eindhoven
|
The Netherlands
|
www.sirius-medical.com
Display Icons
Detection Principle
001118v4.0 US EN
How to calibrate?
Hold the probe vertically in the air and at least 30 centimeters away from any other
equipment. Push the calibrate button and release. Hold the probe still. The calibration
takes approximately 5 seconds
When to calibrate?
Before each new patient and when the system indicates it.
No signal?
Gently palpate (push) the tissue with the probe to reduce the distance between Pintuition
Probe and Seed.
Sterility?
The Pintuition Probe is not a sterile medical device. Always cover with a new, legally
marketed sterile cover before each use in a sterile area.
Note: Numbering not consecutive. Numbers relate to remainder of instructions for use. See Section 3.
SIRIUS PINTUITION DETECTOR
Quick Reference Guide
Warning
! This sheet is not intended as a replacement of the complete Instructions for Use. Please read the complete Instructions for Use booklet
supplied with the Pintuition Detector and Pintuition Probe before first use, or refer to www.sirius-medical.com/support for
the latest version.