4.1
C
ONNECTING THE
P
INTUITION
P
ROBE
To use the Pintuition Detector, the Probe must be connected to the Base Unit:
1.
Align the white arrow on the connector
(16)
with the white arrow on the Probe
connector port
(5)
.
2.
Gently push the connector into the port until a locking click is felt.
Figure 6: Connector
(16)
with white alignment arrow indicated.
If the Probe is correctly connected, the right bottom corner
(11)
of the Base Unit display
(2)
will show a blue check mark and beep A1 is generated (see Section 3.2.6 (Audio)).
If the Probe is not (correctly) connected, the right bottom corner of the Base Unit display
(2)
will show a red ‘X’.
If this occurs, connect, or reconnect the Probe until the display
(2)
shows that the Probe has
been correctly connected in the right bottom corner
(11)
.
If a problem persists, see Section 8 (Troubleshooting).
Figure 7: Display info when the Probe is correctly connected (left) and not (correctly) connected (right).
4.2
C
ALIBRATING THE
P
INTUITION
D
ETECTOR
Caution.
Calibrate the Pintuition Detector before each clinical use and when the
device provides unexpected output.
If the Pintuition Probe
(13)
is correctly connected and the Pintuition Base Unit
(1)
is turned on,
the Pintuition Base Unit
(1)
will require calibration before performing measurements. The
screen will indicate this by showing a red scale icon.
Figure 8: Display info when the Detector requires calibration (left) and when the calibration is in progress (right)
To perform calibration:
1.
Hold the Probe vertically in the air and at least 30 centimeters away from any
electrical equipment or other magnetic disturbances.
2.
Push the calibrate button
(4)
on the Base Unit
(1)
once and release.