User manual - WEIGHT METER SWI-94
”C tYPE”
- calibration type.
Possible values:
“dAtA”
- theoretical calibration („Data Sheet” type calibration) according to data sheet
of load cell transducers (strain bridges),
“rEAL”
- dead weight calibration (entry of values for certain known loads).
„
CALibr
” menu shows option related to selected calibration method only. When
dead weight calibration is selected then
„LoAd”
parameter is visible
.
When
theoretical calibration is selected then parameters „
r LoAd
” and „
r out
” are visible
in place of „
LoAd
”.
”r LoAd”
- Rated Load calibration parameter. This parameter is expressed in
kg
and can be
changed in range from 000000 to 999999, making allowance for „
rESOL
”
parameter.
In case of connection of two, three or four transducers parameter „
r LoAd
” should
be set to arithmetic sum of nominal weights of all the sensors.
”r out”
- Rated Output calibration parameter (output value of load cell transducer at
nominal load). This parameter is expressed in mV/V and can be set in range from
0.0000 to 1.9999 (when
„rAnGE”
= 2mV/V) or from 0.0000 to 3.9999 (when
„rAnGE”
= 4mV/V).
In case of connection of two, three or four transducers parameter „
r out
” should be
set as arithmetic mean of all sensors signals.
”LoAd ”
- function allowing dead weight calibration. After selection of this function it is
necessary to enter some value (expressed in
kg
) and confirm entered value by
pressing [
ENTER
] or press [
ESC
] when value needn't to be changed. To cancel
calibration press [
ESC
] after questions „
rEADY?
” appear. Pressing [
ENTER
] after
„
rEADY?
” appear causes storing current value of input signal. If input signal value
exceeds permissible range then calibration fails and message „
Err
” is displayed.
Weight is expressed in
kg
, and can be changed in range from 000000 to 999999
making allowance for „
rESOL
” parameter.
Due to measurement precision it is recommended to use weight not less than 2/3
of full scale weight while dead weight calibration. Best results can be obtained after
calibration with dead weight equal to maximum expected weight put of the scale.
”S ZEro”
- this function allows to storing momentary value as “zero” of recalculation
characteristic. Storing is done directly after pressing of [
ENTER
] button.
”Z oFFS”
- Zero Offset calibration parameter. This parameter is expressed in
kg
and can be
changed in range from 000000 to 999999, making allowance for „
rESOL
”
parameter.
”FiLtEr”
- this parameter defines input signal filtration rate. It can be changed in range from
0 (lowest filtration) to 5 (highest filtration) which corresponds to
26Hz
,
17Hz
,
8Hz
,
4Hz
,
1Hz
,
0,5Hz
. Bigger filtration level causes slower changes of measurement
results because measurements are realised with longer time interval. While fast
33