3.
TECHNICAL DATA
Power supply voltage
(depending on version)
External Fuse (required)
Power consumption
85...
230
...260V AC/DC; 50 ÷ 60 Hz
or 19...
24
...50V DC; 16V...
24
...35V AC
T - type, max. 2 A
typically 15 VA; max. 20 VA
Display (depending on version)
3.5”, TFT color graphic display, 320 x 240 pixels, with
LED backlight
Sensor power supply output
24V DC ± 5% / max. 200 mA,
Basic communication interfaces
RS 485, 8N1/2, Modbus RTU, 1200 bit/s ÷ 115200 bit/s
USB Host port,
Digital i
nput ( DI )
1 input 0/15..24V DC,
galvanic isolation (low state:
0÷1V, high state:8÷24V)
power consumption
: 7,5 mA / 24V,
isolation: 1min @ 500V DC.
Optional communication hardware
configurations*
Additional Ethernet port with or without IP67
connectors**
Additional RS-485 port + Ethernet port with or without
IP67 connectors**
Optional input hardware
configurations*
2/4x universal inputs with flowmeter mode **
2x counter inputs with tachometer mode**
4x binary inputs**
Optional output hardware
configurations*
2/4x Relay 1A/250V**
2/4x Passive current outputs
**
2/4x OC-type outputs with PWM mode**
2/4x Passive current output + 2/4x Relay module
1A/250V**
2/4x Passive current output + OC-type outputs with
PWM mode**
2
/4x Relay module 1A/250V + 2/4x OC-type outputs
with PWM mode**
Protection level
IP 65
Housing type
Housing material
Wall mounted
ABS + polycarbonate
Housing dimensions (L x h x D)
166 x 161 x 103 mm
166 x 195 x 103 mm
9