Technische Änderungen vorbehalten
4.2
What must be programmed when connecting a flow sensor?
The TD 8250 is set-up for the connected flow sensor.
This is carried out under menu reference
“00 - pulse flow sensor”
, menu reference
“09 -
density”
, and under menu reference
“08”
at the position
“counter input”
.
The pulse volumes for SIKA flow sensors can be obtained from the table.
Type
Material Input
value
Menu-reference 00
Input value
Menu-reference 08
VZ 0,025 Ductile iron / stainless steel
0,025 cm
3
XXX0 2 channels
VZ 0,04
Ductile iron
0,040 cm
3
XXX0 2 channels
VZ 0,2
Ductile iron / stainless steel
0,245 cm
3
XXX0 2 channels
VZ 0,4
Ductile iron
0,400 cm³
XXX0 2 channels
VZ 1
Ductile iron / stainless steel
1,036 cm
3
XXX0 2 channels
VZ 3
Ductile iron / stainless steel
3,000 cm³
XXX0 2 channels
VZ 5
Ductile iron / stainless steel
5,222 cm
3
XXX0 2 channels
VZ 0,2 AL Aluminum
0,245 cm
3
XXX0 2 channels
VZ 2 AL
Aluminum
2,000 cm
3
XXX1 1 channel
The “X” characters in the “Input value Menu reference 08“ column are of no significance for the set-
ting up of the flow sensor.
Example:
An VZ0.04 flow sensor is connected. The pulse volume is 0.040 cm
3
. The VZ 0.04 Flow sensor
has a 2 channel layout.
The value 0.040 is entered under menu reference “00“.
A “0” is entered under menu reference “08” at the position designated “counter input”.
Additionally a density factor can be adjusted to indicate the mass.
Density of medium will be adjusted under menu “
09 – density factor
”. For volumetric
measurement density factor has to be adjusted to “
1.000
”.
The impulse volume (cm³/Impulse) of the SIKA turbine flow sensor is shown on the name
plate and on the enclosed test certificate. The pulse rate refers to the rate of 1 litre of the
medium used in the calibration.
The impulse volume is calculated as follow:
1000 ccm : number of pulses / litre = impulse volume (cm³/Impulse)
for example:
1000 ccm : 9854 Imp/l = 0.1014816... (cm³/Impulse)
0.101 has to be entered under menu reference 00.