3.6
Memory functions
The MC 75.2 is capable of storing 10,000 values in one or more acquisition bursts.
•
Press the key F2 to open the ‘IN’ window and display
the
Measurement
menu bar.
•
Open the Configuration menu by pressing the key
F1.
•
Select the
Menu
function and then
Memory
.
•
Confirm with ENTER again.
Two new functions,
Memory
and
MREC1
, then appear in the bottom bar (replacing the func-
tions Rest stat. And Hold).
The left
and right
navigation keys can be used to switch from one mode to the other.
•
Pressing the function key
F4
(MREC1) stores the
current acquisition.
•
Pressing the key
F3
(Memory) gives access to all the
memory functions.
RUN:
Starts the storage of data as configured using the
Pa-
rameters
function. The icon is shown in the Measure-
ments window.
STOP:
Stops the current storage operation.
PARAMETERS:
This can be used to define::
- Burst size (10,000 values max.).
- Sampling period from 0.5 sec. to 30 min.
- Type of trigger (none, low level, high level).
If high-level or low-level trigger is selected, the
trigger level and number of data points to be
stored after this trigger must be defined.