9 Technical data
The technical data of customised versions may differ from the data in these instructions.
Please observe the information specified on the type plate.
9.1 Characteristics VMZ.2
Type
VMZ03
VMZ06
VMZ08
VMZ15
VMZ20
VMZ25
Measurement device characteristics
Flow range [l/min]
0.1…2
0.25..5
1…20
2.5…50
5…200
12.5…250
Accuracy*
±0.7 % of reading ±0.3 % of range
Repeatability
±1%
Output signal starting
from [l/min]
0.05
0.1
0.25
1
4
5
Max. flow rate [l/min]
2.5
6
25
60
240
300
Response time
< 100 ms
Indication
LED green, flow proportional flashing
Output signal characteristics
Frequency output:
- Pulse rate / K-Factor
[pulses/l]
10 000
4000
1000
400
200
80
- Resolution [ml/pulse]
0.1
0.25
1.0
2.5
5.0
12.5
- Signal shape
Square wave signal • duty cycle 50:50
Push-Pull • NPN open collector • PNP open collector
- Signal current
≤
100 mA
Analogue output 4...20 mA (optional)
Corresponds to flow rate
[l/min]
0...2
0...5
0…20
0…50
0…200
0…250
Max. burden
250 Ω
to GND
Analogue output 0.5...10 V (optional)
Corresponds to flow rate
[l/min]
0...2
0...5
0…20
0…50
0…200
0…250
Electrical characteristics
Supply voltage
12...24 V
DC
(±10 %)**
Power consumption
Max 3.6 W
Electrical protection
measures
short-circuit proof • protected against polarity reversal
Electrical connection
4-pin plug M12x1
Degree of protection
(EN 60529)
IP 65 (only with a connected coupling)
* test conditions: Water 23 °C
** Voltage output 0.5…10 V only available with 16…24 VDC