Instruction Manual ScrubberGuard
Servicing
98
14830E/2
WORKSTEP
ADDITIONAL INFO / IMAGES
8.
Move the slide holder upwards, rinse off the
pH sensor with distilled water and pat dry.
9.
Carrying out recalibration for nominal
value 2:
Press the arrow button at the bottom right to
switch to
Nominal value 2
and repeat steps
3 to 8 with the second calibration solution.
10. Mount the pH sensor in the holder according
to Section 4.3.1.
9.3.3
Replacing pH sensors that have been configured by SIGRIST
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the old pH sensor from the holder ac-
cording to Section 4.3.1.
2.
Mount the new pH sensor according to Sec-
tion 4.3.1.
3.
Put the system back into operation.
9.3.4
Installing an unconfigured pH sensor
This process is only applicable if a new sensor has not been ordered from SIGRIST-
PHOTOMETER.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
In the
Local ……
menu, access the
Digi. interf.
submenu.
2.
Remove the old pH sensor from the holder ac-
cording to Section 4.3.1.
3.
Mount the new pH sensor in the holder and
connect according to Section 4.3.1.