Instruction Manual ScrubberGuard
Settings
66
14830E/2
8.7
Back up configured data
These measures can be of use to the service engineers for service purposes.
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Local
…….
button.
4.
Press the
System info.
button.
If the desired menu does not
appear, press the arrow key bot-
tom right.
5.
In the
User -> SD
and
Expert -> SD
sub-
menus press the
Copy
function.
The user and expert data are cop-
ied to the microSD card. After a
successfully completed procedure,
acknowledge with the
OK
button.
6.
Press the
Meas
button.
The instrument is in measuring
mode again.