Operation
Instruction Manual ScrubberGuard
14830E/2
47
7
Operation
7.1
Control elements of the ScrubberGuard on the cabinet
The ScrubberGuard is controlled via four control elements located on the cabinet. These are
three buttons plus a control input for external actuation. Only one control element may be
active at any one time. Activation is confirmed by the respective button lighting up. If mul-
tiple functions are activated, this leads to a fault that is signaled by the flashing buttons. To
rectify this problem, all functions must be deactivated.
▪
In order to ensure correct operation, the AquaScat SG (A) must be connected and the
communication between the AquaScat SG (A) and cabinet must be established.
▪
If the optional inlet pump is installed, the supply line must be filled with water before-
hand.
Main switch
: The main switch is used to
switch the system on and off.
Figure 17: Control elements on the Scrub-
berGuard
The pump may only be operated
dry for a maximum of 30 seconds.
Danger of overheating!
Remote Signal
(signal lamp): Mea-
surement can be controlled by the con-
trol system via a digital control input or
field bus. To do this,
External
must be
set in the
Scrubber\Scrubber\Control
menu and a connection to the control
system must be established. The control
signal has the same function as the Mea-
surement ON/OFF button. The signal
lamp indicates the current state of the
control signal in the control system.
Measurement
ON/OFF
button: Pressing
this button starts the water circulation.
The inlet valve is opened and the internal
circulation is started. After the maximum
tank level has been reached, the return
valve opens and the return pump starts.
To stop the water circulation, press the
button again.
Loop/Spiking
button: Pressing this button starts the spike mode. This is used for
checking the calibration of the instruments using a calibration solution. Spike
mode activates the internal circulation of the measuring instruments. The sample
supply and return lines are closed. To cancel this mode, press the button again.
Drain tank
button: Pressing and holding this button drains the system. To cancel