Initial start-up
Instruction Manual PhaseGuard C/ T/ HT
11026E/4
17
6.
Initial start-up
6.1.
Commissioning procedure
WORKSTEP
PROCEDURE
1.
Check mounting and in-
stallation.
Ensure that the photometer and components are correct-
ly mounted and connected.
2.
Connect power supply to
the photometer.
Section 5
3.
Connect photometer to
PC.
3.1: Remove cover (arrow) of the USB interface.
Damage to the photometer caused by
moisture inside the instrument:
The cover to the USB interface may be removed only un-
der dry conditions.
3.2: Connect the PhaseGuard via USB cable to the PC.
The PhaseGuard registers itself as a removable disk in the
Windows operating system.
3.3: Depending on the Windows configuration, a win-
dow opens automatically with the removable removal
disk or you have to open the removal disk manually using
Windows Explorer.
The following files appear:
Info.txt
: information data about the PhaseGuard, e.g.
serial number, software version, hours of operation.
Config.txt
: configuration file.
Zero.txt
: file with which a zero calibration can be
initiated.
4.
Configure.
The following functions
can be set as described in
4.1 to 4.4:
4.1: Open the
Config.txt
file.
4.2: Change the desired parameter. Enter number after
“
=
“.
e.g. 0 for German or 1 for English