Settings
Instruction Manual OilGuard 2 Ex
13003E/3
57
8.3
Setting the limits
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Limits
button.
If the desired menu does not
appear, press the arrow at the bot-
tom right.
4.
Select
Limit 1/2
.
5.
Define
Mode
.
The following selection is available:
Inactive
(limit monitor of this
channel is deactivated)
Exceeded
(limit active when
the set threshold value is ex-
ceeded)
Undershot.
(limit active when
the set threshold value is under-
shot)
6.
Define the upper limit, lower limit, cut-in delay
and cut-out delay with number pad.
Pressing the current number
value takes you to the entry mode.
7.
Press the
Meas
button.
The instrument is in measuring op-
eration again.
The limits have to be configured accordingly so that they are not only displayed, but that the
outputs are also switched.