
Instruction Manual FireGuard 2
Settings
68
13542E/4
Pos : 146 /Bedi enung/FireGuard 2/ Str omausgänge ei nst ellen Fir eG uar d 2 @ 4 \ mod_1442069998235_20. doc x @ 21637 @ 2 @ 1
8.8
Setting the current outputs
This setting only has to be carried out if the optional StromRel module is used.
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Curr. outputs
button.
4.
Select between
C 1 .. 2
.
5.
Select the
Source
menu.
For C1: C1 Turb
For C2: C3 Temp
6.
Select the
Range
menu.
For C1: MR1 0 .. 50 mE/m
For C2: MR2 -20 .. 80 °C
7.
Press the
Meas
button.
The instrument is in measuring
mode again.
Measuring range no. Measuring range
(standard)
Measuring range
(customer-specific)
MR1
0 .. 50
MR2
-20 .. 80
MR3
0 .. 0
MR4
0 .. 0
MR5
0 .. 0
MR6
0 .. 0
MR7
0 .. 0
MR8
0 .. 0
If other measuring ranges are needed, the table above can be reprogrammed as required.
Other parameters for the current outputs are described in the Reference Handbook.
Pos : 147 /---Seit enumbruc h--- @ 0\ mod_1341482862121_0. doc x @ 4073 @ @ 1
Содержание FireGuard 2 Integral
Страница 3: ......
Страница 6: ...Instruction Manual FireGuard 2 Contents 6 13542E 4 This page is intentionally blank ...
Страница 101: ......