Instruction Manual FireGuard 2
Electrical installation
30
13542E/4
Pos : 68 /I nst allation, Elektrisch/FireGuard 2/Ansc hliess en des SI PORT 2 @ 4\ mod_1436441464707_20. doc x @ 21234 @ 2 @ 1
5.2
Connecting the SIPORT 2
DANGER!
Life-threatening voltage due to accidentally released voltage-carrying wires.
The cable glands must be adjusted according to the outer diameter of the cables. The fol-
lowing cable glands are available:
2 x 8 .. 17 mm
3 x 8 .. 13 mm
2 x 5 .. 10 mm
Figure 11: SIPORT 2 when open
Photometer connection
(cable cross-section: 0.2 .. 4.0
mm
2
)
optional version with con-
nector
Mains voltage connection
(cable cross-section: 0.2 .. 4.0 mm
2
)
Fine-wire fuses
(microfuse 250V 2AT RM5)
The SIPORT 2 is always fitted with one
of the following interface modules:
Profibus DP
Modbus RTU
StromRel
Profinet IO
(cable cross-section: 0.08 .. 1.5 mm
2
)
Connection of portable control unit
on SIPORT 2 without housing
Connection of portable control unit on
SIPORT 2 with housing