Instruction Manual DualScat Ex
10118E/5
33
5.6.1
Preparations for recalibration with the checking unit
Make sure to use the proper SIGRIST checking unit: the serial number printed
on the calibration unit must match with the instrument serial number!
Figure 16: SIGRIST checking unit.
Action
1.
Make sure the product pipe is empty, dis-
connect the power supply to the control unit,
and remove the photometer from the pipe.
2.
Clean the sensor head as directed in
Section 5.4.
3.
Remove the calibration unit
supplied with the
instrument
from its box. Clean the calibration
unit, inside and out, with a soft, lint-free rag.
Make sure the serial
number printed on the
calibration unit matches
with the instrument serial
number.
4.
Set the checking unit vertically on a flat sur-
face. Align the photometer so that the
groove on the sensor head mates with the
guide pin on the checking unit.
Be very careful not to insert the pho-
tometer into the checking unit turned
incorrectly or to rotate it during inser-
tion, because this would damage the
solid reference!
Preparations for
checking unit
Base
Filling nozzle
Outlet nozzle
Funnel
Level indicator
Solid reference