Instruction Manual DualScat Ex
24
10118E/5
Action
Display (Example)
Remarks
11.
Set upper threshold:
/
change number
/
change place
Upper limit1
> ... <
Now set the value at
which the limit should
switch on.
12.
> Upper limit1 <
...
Confirm entry.
13.
> Lower limit1 <
0.900 NTU
14.
Lower limit1
> 0.900 NTU <
Activate editing mode.
15.
Set lower threshold:
/
change number
/
change place
Lower limit1
> ... <
Now set the value at
which the limit should
switch off.
16.
> Lower limit1 <
...
Confirm entry.
17.
> LI2 source <
Meas.value25
Same procedure for 2
nd
limit
steps 6 to 16.
18.
+
(simultaneous)
90° 2.23 NTU 5
Instrument in normal oper-
ation.
4.9
Setting the access code
You can set your own access code to protect your DualScat Ex settings
against unauthorized manipulation.
Action
Display (Example)
Remarks
1.
Access code
> ▓00000 <
If no individual access
code was set, proceed to
step 3.
2.
Enter code:
/
change number
/
change place
Access code
> ...... <
Enter your own access
code here.
3.
* SENSOR *
* CHECK *
Instrument in service
mode.
4.
4 x
* CONFIGURATION*
5.
3 x
> Access code <
000000
6.
Access code
> 000000 <
Activate editing mode.
7.
Enter new code:
/
change number
/
change place
Access code
> ...... <
Enter new code below to
make sure you don't for-
get it!
Access code
Содержание DualScat Ex
Страница 2: ... SIGRIST PHOTOMETER AG Subject to change without notice 9 2016 ...
Страница 6: ...Instruction Manual DualScat Ex iv 10118E 5 ...
Страница 53: ...Instruction Manual DualScat Ex 10118E 5 47 ...
Страница 54: ...Instruction Manual DualScat Ex 48 10118E 5 11 Appendix ...
Страница 55: ...Instruction Manual DualScat Ex 10118E 5 49 ...