Instruction Manual DualScat Ex
10118E/5
15
3.3
Initial start-up
For the initial start-up, work through the following table. If you run into trou-
ble, consult Section 6.
Action
Remarks
1.
Check to make sure the photometer
and control unit are properly in-
stalled and connected.
Sections 3.1 and 3.2.
2.
Activate the power supply to the
control unit.
After a few seconds, either one or
two readings appear on the control
unit display (depending on instru-
ment version).
3.
Set the language spoken in your re-
gion.
Section 4.5
Now the menu texts will appear in
the desired language.
4.
Run through a sensor check.
Section 6.3
If no fault message has appeared,
the photometer is ready to start
measuring.
5.
Set the measuring range required for
your duty conditions.
Section 4.6
If automatic range switching is de-
sired (standard), you can skip this
step.
6.
Set the relay functions to suit your
measuring duty.
Section 4.7
7.
Protect your settings against unau-
thorized manipulation with an ac-
cess code.
Section 4.9
If you don't need an access code,
skip this step.
Initial start-up