WORKSTEP
ADDITIONAL INFO / IMAGES
8.
Move the slide holder upwards, rinse off the
pH sensor with distilled water and pat dry.
9.
Carrying out recalibration for nominal val-
ue 2:
Press the arrow button at the bottom right to
switch to
Nominal value 2
and repeat steps 3
to 8 with the second calibration solution.
10. Mount the pH sensor in the holder according
to Section 4.3.1.
9.3.3
Replacing pH sensors that have been configured by SIGRIST
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the old pH sensor from the holder ac-
cording to Section 4.3.1.
2.
Mount the new pH sensor according to Section
4.3.1.
3.
Put the system back into operation.
9.3.4
Installing an unconfigured pH sensor
This process is only applicable if a new sensor has not been ordered from SIGRIST-
PHOTOMETER.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
In the
Local ……
menu, access the
Digi. interf.
submenu.
2.
Remove the old pH sensor from the holder ac-
cording to Section 4.3.1.
3.
Mount the new pH sensor in the holder and
connect according to Section 4.3.1.