Servicing duties on the pH sensor
9.3.1
General handling of the pH sensor
9.3.1.1
Overview with pH sensor
The calibration process is designed for use with Hamilton calibration solutions (500 ml con-
tainer). Although it is possible to use other calibration solutions, SIGRIST-PHOTOMETER ex-
pressly recommends using the Hamilton standards. Recalibration is made at two points (pH 4
and pH 7).
Figure 29: Overview of calibration
Complete pH sensor
a: Connection/electronic system
b: Holder
c: Shaft (electrode)
d: Measuring tip
Connector
Slide holder
Bottle with calibration solution