Instruction Manual ScrubberGuard
Troubleshooting
104
14830E/1
10
Troubleshooting
10.1
Pinpointing faults
DETECTABLE FAULT
MEASURE
No reading
Check whether the supply voltage is connected.
Check whether the fine-wire fuse is OK. Section 10.5
Fault message in the display
Analyze the fault message according to Section 10.2
to Section 10.4.
The reading appears to be
wrong
Ensure that the sample to be measured corresponds
to the operating conditions. Section 2.4
Ensure that the sample is free from air bubbles. Sec-
tion 9.4.2
Carry out recalibration. Section 9.4.3
Check whether the system is correctly mounted. Sec-
tion 4
Ensure that the servicing duties have been performed
according to the servicing schedule. Section 9
Table 2: Pinpointing malfunctions
If the listed measures do not lead to the desired results, please consult customer service. Sec-
tion 11