Instruction Manual AquaScat S
Troubleshooting
98
13798E/3
FAULT MESSAGE
DESCRIPTION
POSSIBLE CAUSES
POWERBOX
Actuation of the power box
has been disturbed.
Connection to the power box
has been interrupted.
IO PORT
The connection between the
NG_Haupt and NG_Bedi print
in the SICON has been dis-
turbed.
Cable disconnected.
Plug connection defective.
HUMIDITY
The relative humidity in the
instrument has risen above
50%.
The desiccant is saturated.
The gaskets on the electronic
component are defective.
The instrument was open too
long.
Table 4: Possible fault messages
10.4
Prioritized fault messages and their effect on operation
CAUTION!
When there is a prioritized fault, the cause of the malfunction is serious.
PRIO (PRIORITIZED FAULT)
If a prioritized fault occurs during operation, it has the
following effects:
The measuring values go to 0.
Prioritized faults can be cleared only by a service
engineer.
When the
Prio
message occurs,
the color of the
status display changes to
red
and the text de-
scribes the prioritized fault in question.
Example:
PRIO DEFAULT VALUES