Mounting
Instruction Manual AquaScat S
13798E/3
27
4.4
Mounting the extractable assembly
4.4.1
Extractable assembly: Mounting the probe in a vertical pipe
A 2" threaded coupling (pos. 2) must be welded onto the pipe beforehand according to the
AquaScat_S_WA-MB
drawing.
Before mounting the extractable assembly, remove the probe as detailed in Section 9.5(step
1 to 8). The extractable assembly must be sealed on the 2" threaded coupling and posi-
tioned so that the handle of the shut-off valve (pos. 3) is at the top. The probe can then be
reinstalled as detailed in Section 9.5(step 10 to 17).
Insert the outlet hose (pos. X) in the hose coupling and connect the end of the hose with the
drain.
Figure 8: Overview of a measuring point with extractable assembly
AquaScat S (absorber)
2" threaded coupling
Shut-off valve
Marking on the rating plate (shows the
position of the angled edge (pos. 9))
Handle
Hose for pressure compensation
Tri-Clamp
Changeover valve
The angled edge of the sensor
head must be positioned against
the flow direction.
Flow direction of the sample