Operation
INSTRUCTION MANUAL AquaScat 2
11778E/1
33
7.2.3.
Functions of the log screen (Log button)
The screen logger works independently of the data logger, which is set in the
Logger
menu
and writes to the microSD card.
The screen logger records the data of the last 32 days in one minute intervals. The data can
be called up from the Log menu.
If the instrument is out of operation for more than 32 days, the logger data is restarted. An
hour glass is shown for about 1.5 minutes in the graphic display. During this time no logger
data is available.
The
Log
button is in the graphic screen only after the
Diag
button has been pressed. When
the
Log
button is pressed, the following screen appears:
Figure 9: Functions of the Log display
Cursor shows the time position which
is represented at pos. 4.
The cursor position can be changed
either by briefly touching with your
fingertip or by pressing the </> but-
tons.
Represented time period
The following time ranges can be
set: 3 min. / 15 min. / 1 hr. / 3 hr. /
9 hr. / 1 day / 3 days / 10 days /
32 days
Indicates how much of the total time
period is currently represented.
Measuring value which was meas-
ured at the cursor position.
</>
: Moves the cursor position. The cursor moves faster when these buttons are held
down longer.
<</>>
: Jumps forward or backward by the time period set in point 2.
-/+
: Increases (+) or decreases (-) the screen section around the cursor position.
In the
Display\General\Values
menu (
Reference Manual) you can define whether mini-
mum, maximum or average values are to be displayed.
Pressing the Diag button takes you to the graphical representation.