![Signametrics Corporation SMU2060 Скачать руководство пользователя страница 123](http://html1.mh-extra.com/html/signametrics-corporation/smu2060/smu2060_operators-manual_1268488123.webp)
123
Signametrics
Description
Set the measurement delay of AC based Capacitance.
#include "SMU2060.h"
#include "USBDMMUser.h"
int
DMMSetACCapsDelay(int
nDmm
, double
ldDelay
)
Remarks
This function sets the AC based capacitance measurement delay, which is the time the
measurement system settles. The default value is 0s. This function can set this value
from 0.0 to 10.0 seconds. Since the DMM is optimized for the defalut value, it is possible
that changing this value will introduce additional error. A negative delay value causes the
range limits to be disabled, making it possible to measure values greater than defined by
the selected range. For instance a value of -0.01 will set the delay to 0.01s and remove
the range limit. The default is hat range limits are enabled.
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
ldDelay
double
The time the DMM is allowed to settle the measurement.
Can be set beetween -10.0 and 10.0 seconds. A negative value
disables range limits.
Return Value
Integer error code.
Value Meaning
DMM_OKAY
Operation successfully completed.
Negative Value
Error code
Example
DMMSetACCapsDelay(0, 0.25); // Set measurement delay to 0.25s
DMMSetACCapsLevel
SMU2060
SMU2064
Description
Set the level of the AC voltage source for a peak value during In-Circuit caps.
#include "SMU2060.h"
#include "USBDMMUser.h"
int
DMMSetACCapsLevel(int
nDmm
, double
ldVolts
)
Remarks
This function sets the AC peak voltage level for the In-Circuit Capacitance measurement
function. This value is used on any of the AC Caps calibration and measurement.
Following setting of this function, it is necessary to perform open calibration of the AC
Capacitance ranges to be used. Since the DMM is optimized for the default value, it is
recommended not to use this function, maintaining the default 0.45V peak value.
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
ldVolts
double
Peak value of AC voltage to be set. Can be 0.1V to 5.0V
Return Value
Integer error code.
Содержание SMU2060
Страница 26: ...Signametrics 26 PXI Instrumentation Switching modules SMX4030 SMX4032 IVI COM driver...
Страница 81: ...81 Signametrics Postive Value 100 Warning code Example char cBuf 64 int status status DMMGetCalDate 0 cBuf...
Страница 106: ...Signametrics 106 Postive Value 100 Warning code Example char cBuf 64 int status status DMMPeriodStr 0 cBuf...