Signametrics
120
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
lpszReading
LPSTR
Points to a buffer (at leaset 64 characters long) to hold the
converted results. The return value will consist of a leading sign, a
floating-point value in exponential notation, and a untis specififer.
Return Value
The return value is one of the following constants.
Value Meaning
Negative Value
Error code
Positive value < 100
The length of the returned string.
Positive value > 100
Warning code
Example
char cBuf[64];
int status = DMMReadStr(0, cBuf);
DMMReadTestV
SMU2060
SMU2064
Description
Return the exact voltage applied during Leakage test.
#include "SMU2060.h"
int DMMReadTestV
(
int
nDmm,
double
*
lpdTestV
)
Remarks
This is function requires the DMM to be in LEAKAGE measurement. It measres the
voltage applied to the device during leakage measurement, placing the double-precision
floating-point result in a location pointed to by
lpdTestV
. The starndard measurement
functions,
DMMRead()
,
DMMReadStr()
and
DMMReadNorm()
return the value of
the leakage.
Parameter Type/Description
nDmm
int
Identifies the DMM. DMMs are numbered starting with zero.
lpdTestV
double *
Points to the location which holds the result.
Return Value
The return value is one of the following constants.
Value Meaning
DMM_OKAY
Operation successfully completed.
Positive Value
Warning code
Negative Value
Error code
Example
double Vt; int state; state = DMMReadTestV(0, &Vt);
Содержание SMU2060
Страница 26: ...Signametrics 26 PXI Instrumentation Switching modules SMX4030 SMX4032 IVI COM driver...
Страница 81: ...81 Signametrics Postive Value 100 Warning code Example char cBuf 64 int status status DMMGetCalDate 0 cBuf...
Страница 106: ...Signametrics 106 Postive Value 100 Warning code Example char cBuf 64 int status status DMMPeriodStr 0 cBuf...