Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
3 Safety
Version 03/2014, Rev. 2.11 of 11/08/2017
• sb
21 / 79
Translation of the original operating manual, part no. 0700902
Pos : 56 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0060-0070 Sicherheit von R ot oren und Zubehör @ 44\ mod_1470119200568_68.doc x @ 437008 @ 3 @ 1
3.6.7
Safety of rotors and accessories
Pos : 57 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0060-0070- 0010 Lebens dauer @ 26\ mod_1405319032612_68. doc x @ 192143 @ 3 @ 1
3.6.7.1
Service life
The rotors and accessories have a limited service life.
WARNING
• Perform regular checks (at least once per month) for safety reasons!
• Pay special attention to changes, such as corrosion, cracks, material
abrasion, etc.
• After 10 years, they must be inspected by the manufacturer.
• After 50,000 cycles, the rotor must be scrapped for reasons of safety.
• If other data concerning the service life are engraved on the rotor or
bucket, these data shall apply accordingly. For example, a bucket with
the engraving “max. cycles = 10,000” has a service life of 10,000
cycles, and a rotor with the engraving “Exp. date 02/20” must be
scrapped in February 2020 at the latest (see figure).
Fig. 4: Different service life
– engraving on the bucket/rotor
NOTE
• Refer to the table of rotors and accessories with a different service
life (see chapter 11.4 - "Table of the service life of rotors and
accessories")!
Pos : 58 / 010 U ni vers al module/ Seitenwec hsel @ 0\ mod_1202116244312_0. doc x @ 105 @ @ 1