The large numbers in the middle of the screen show conductivity value tested under
the temperature of +20
℃
or true conductivity values of camber specimen by corrected
coefficient η(φ) compensated.
c
.
η(φ)=□□□□ is the curvature corrected coefficient of the camber, it is only effective
after the camber compensation opened.
d
.
MS/m or %IACS is unit for conductivity value, or Ω·mm²/m is the unit for the
resistivity (units setting refer to 5.3.2.4).
e
.
σ(Tc)=□□□□of the bottom screen said conductivity values which measured under
temperature in bracket .σ(Tc)shows temperature value in brackets is compensation
temperature value, its equal to the T
A
temperature value get after four homes into the
rounding.
f
.
T
A
=□□□
℃
indicates the current temperature of testing pieces measured by a
temperature sensor or manual input, or display T
A
=None under pseudo temperature
compensation mode (setting method see 5.3.4.3 temperature compensation mode).
g
.
f=
□□□
KHz indicates the working frequency (setting method see5.3.4.6
frequency and probe selection)
h
.
a
0
=□□□□□ indicate the temperature coefficient of testing pieces (a=None is invalid
under pseudo temperature compensation mode or continuity measurement interface)
i
.
<Continuity measurement >interface, the icons
on the screen said alarm beeps
open, no icon is closed. σ
H
=□□□
,
σ
L
=□□□ said conductivity high values and low
values of separation point ( seeing 5.3.4.2 measurement interface mode).
5.3.2.4 Measurement unit
Measuring interface has two units of selecting methods, one is default instrument unit "MS/m
and % IACS" (see 5.3.4.7 unit selection), which used in data storage, calibration, measurement
mode. Another is quick convert method, in the measurement of interface, press "
UNIT
DEL
" key,
press key once will converse the unit once, such as "MS/m
、
%IACS
、
Ω·mm²/m", shut down and
then open or from other interface returns to measurement interface, measurement unit recovery to
the original default units.
5.3.2.5 Methods of compensation temperature
a.
When the specimen current temperature TA=□□□
℃
sets as reading temperature of
temperature sensor (automatic temperature compensation for temperature sensor, see
5.3.4.3) , the temperature sensing end is close to the surface of the specimen until
temperature stable, press "
η(φ)
T
A
" button to select TA value(TA display block cursor), press
the "OK" button, σ(Tc)values in brackets immediately modified into TA temperature value
and stored until the next change. Conductivity value
(
at 20
℃)
displayed in the middle
screen will update according to the temperature. Remark: When T
A
=???, compensation
temperature value T
C
can’t be modified.