8 Maintenance and service
Version 10/2012, Rev. 1.22 of 08/05/2020
• sb
65 / 91
Translation of the original operating manual, part no. 0701802
Pos : 187 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/080 War tung und I nst andhaltung/ 080-0010- 0040 Tragbolz en (außer 2-6, 2-6E, 2-7, 2-16P, 2- 16KL) @ 28\m od_1405320624792_68. docx @ 194915 @ 3 @ 1
→
8.1.4
Load bearing bolts
Only greased load-bearing bolts ensure a uniform swing-out of the buckets
and, therefore, the smooth operation of the centrifuge. Load-bearing bolts
that are insufficiently greased may cause the centrifuge to stop due to an
imbalance.
• Clean the load-bearing bolts and bucket groove in order to remove the
old grease.
• Apply a small amount of heavy-duty grease for load-bearing bolts (ref.
no. 71401, see the following picture) to both load-bearing bolts of a
bucket.
Fig. 33: Sufficient quantity of grease for one bolt
• Install the bucket and swing it manually back and forth once in order to
distribute the grease.
• Repeat this process with all the other buckets.