![Sigma 147161 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/sigma/147161/147161_operating-manual_1267884029.webp)
Refrigerated Centrifuge with Heating System Sigma 4-16KHS
6 Using the centrifuge
Version 10/2012, Rev. 1.22 of 08/05/2020
• sb
29 / 91
Translation of the original operating manual, part no. 0701802
Pos : 102 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060 Betrieb========================================= @ 27\mod_1405319198895_68. docx @ 192423 @ 1 @ 1
→
6
Using the centrifuge
Pos : 103 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0010 Erst e Inbetriebnahme @ 27\m od_1405319200130_68. docx @ 192437 @ 2 @ 1
→
6.1
Initial start-up
DANGER
• Before the initial start-up, please ensure that your centrifuge is
properly set up and installed (see chapter 5 - "Set-up and
connection").
Pos : 104 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 105 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020 Ei nsc halt en (auß er 6- 16HS) @ 27\mod_1405319201340_68. docx @ 192451 @ 2 @ 1
→
6.2
Switching the centrifuge on
• Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
Pos : 106 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 107 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0010 Ö ff nen und Sc hließ en des Dec kels - 1 Sc hloss @ 27\m od_1405319202508_68.docx @ 192465 @ 3 @ 1
→
6.2.1
Opening and closing the lid
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
• Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
• To close, press with both hands slightly on the lid until the lock is
audibly locked.
WARNING
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!
Pos : 108 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 109 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0019 Ei ns etz en v on Rotor en und Zubehör ( auß er 8K)-------------------------------- @ 27\m od_1405319204492_68.docx @ 192493 @ 3 @ 1
→
6.2.2
Installation of rotors and accessories
Pos : 110 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 111 /100 Sigma/110 BA Z entrif ugen Sigma (Proj ekt e)/ 4-16KS_4- 16KHS/060 Betri eb/060-0020-0020- 0010 Ei nsetzen ei nes Rotors 4- 16KS_KHS @ 29\mod_1405326945297_68. docx @ 198765 @ 4 @ 1
→
6.2.2.1
Installation of a rotor
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Lower the rotor with its central bore straight down onto the motor shaft.
• Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 10 Nm (Option: torque wrench 10 Nm, part no. 17060). In
doing so, hold the rotor at its outer rim.
• Follow the safety instructions and hazard warnings (see chapter 3 -