5 Set-up and connection
32 / 113
Version 11/2017, Rev. 1.7.1 of 08/12/2022
• sb
Translation of the original operating manual,
The centrifuges are equipped with a mains power switch with an integrated
thermal circuit breaker.
• Switch the unit off by actuating the mains power switch.
• If it has tripped, let the circuit breaker cool for approximately 2 minutes.
• Switch the unit on.
The centrifuge is now ready for operation.
→
5.2.2
Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp B fuses that are to
be provided by the customer.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.