Sigma Space MiniMPL-532-C Скачать руководство пользователя страница 1

 

 

Mini Micro Pulse LiDAR System 

MiniMPL-532-C  

 

Operations Manual 

 

Version: August 2016 

Содержание MiniMPL-532-C

Страница 1: ...Mini Micro Pulse LiDAR System MiniMPL 532 C Operations Manual Version August 2016...

Страница 2: ...MiniMPL 532 C Operations Manual Sigma Space Corporation 2016 2 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 3: ...rchase of the Mini Micro Pulse LiDAR System from Sigma Space Corporation Please fill in the following system information for your records This information may be requested by Sigma Space for obtaining...

Страница 4: ...MiniMPL 532 C Operations Manual Sigma Space Corporation 2016 4 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 5: ...ACKING AND INSTALLATION 11 2 1 Power Requirements at MiniMPL Site 11 2 2 MiniMPL Site Environment 11 3 HARDWARE SETUP 12 3 1 MiniMPL Placement 12 3 2 Data Acquisition Computer Connections 13 4 MINIMPL...

Страница 6: ...MiniMPL 532 C Operations Manual Sigma Space Corporation 2016 6 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 7: ...ed representative Mechanical Optical and Environmental Safety M1 The MiniMPL is not weatherproof and must not be exposed to rain or excessive humidity levels above 80 The operating and storage tempera...

Страница 8: ...4600 Forbes Boulevard Lanham MD 20706 United States of America www micropulselidar com Model No Serial No Manufacture Date Made in U S A This equipment conforms to provisions of US 21 CFR 1040 10 11 a...

Страница 9: ...MiniMPL 532 C Operations Manual Sigma Space Corporation 2016 9 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 10: ...collected and displayed in real time on the data acquisition computer The optical transceiver is integrated with the electronic system in a single box to maximize the portability of the unit The data...

Страница 11: ...g or bumping the MiniMPL may cause serious damage to the components inside The MiniMPL may be stored in its original shipping container when not in use between 10qC 50qF and 35qC 95qF in a location wi...

Страница 12: ...he front and rear panel views of the MiniMPL instrument Figure 2 MiniMPL Front Left and Rear Right Panel Views 3 1 MiniMPL Placement 1 Always use the attached handle when moving the MinIMPL 2 Place th...

Страница 13: ...ld create disruptions in data transmission 4 MiniMPL Operation Procedures The MiniMPL should be located with an unobstructed view of the atmosphere and the aperture lid should be opened carefully to a...

Страница 14: ...before collecting data To access configuration options press Ctrl K L This will enable the Collection Parameters to be adjusted in the Hardware tab To configure the software for MiniMPL select Config...

Страница 15: ...MiniMPL system should be shut down in the following sequence 1 Click the Stop Collecting button in the SigmaMPL program to stop data acquisition 2 Exit the SigmaMPL program 3 Turn off the KEY SWITCH...

Страница 16: ...532 nm Nd YAG Pulse Repetition Frequency 2 500 Hz Pulse Energy 3 4 J Laser Product Compliance ANSI Z136 1 2000 IEC EN60825 USFDA CDRH 21 CFR 1040 10 11 Receiver Telescope Type Galilean Telescope Focal...

Страница 17: ...trument to the Sigma Space facility for warranty repairs It is recommended that the buyer insure the instrument at their expense Sigma Space may levy a reasonable assessment fee for evaluating the pro...

Отзывы: