30 SSG5000A Programming Guide
System Preset (:SYSTem:PRESet)
Command Format
:SYSTem:PRESet
Instruction
According to the preset type, preset the parameter configuration of the
machine.
Parameter Type
None
Return
None
Default
None
Menu
Utility > Preset
Example
For example, preset signal generator to default configuration:
:SYSTem:PRESet:TYPE DFT
:SYSTem:PRES
Or preset signal generator to current configuration:
:SYSTem:PRESet:TYPE USER
:SYSTem:PRESet:SAVE
:SYSTem:PRES
Or preset signal generator to configuration saved in an existing xml file:
:SYSTem:PRESet:TYPE USER
:SYSTem:PRESet:PATH “Local/test.xml”
:SYSTem:PRES
Preset Save (:SYSTem:PRESet:SAVE)
Command Format
:SYSTem:PRESet:SAVE
Instruction
Save status for preset when preset type is user
Parameter Type
None
Return
None
Default
None
Menu
Utility > Preset
Example
:SYSTem:PRESet:SAVE