SSG5000A Programming Guide 29
Default
None
Menu
Utility > Setting > 10M Adjustment > Reset to Default
Example
:SYSTem:REF:DAC:DEFault
GPIB Address (SYSTem:GPIB)
Command Format
SYSTem:GPIB <value>
SYSTem:GPIB?
Instruction
Set GPIB address of the signal source.
Get GPIB address of the signal source.
Parameter Type
Integer
Parameter Range
1 ~ 30
Return
Integer
Default
18
Menu
Utility > Interface > GPIB Address
Example
SYSTem:GPIB 10
SYSTem:GPIB?
3.3
Preset Subsystem
Preset (:SOURce:PRESet)
Command Format
:SOURce:PRESet
Instruction
Presets all parameters which are related to the selected signal path
Parameter Type
None
Return
None
Default
None
Menu
None
Example
SOUR:PRES