SSG5000A RF Signal Generator User Manua
l
SSG5000A User Manual 80
At this time, the
[RF OUTPUT 50Ω]
connector outputs an RF signal with frequency of 3
GHz and amplitude of 0 dBm.
3.2 Correcting 10MHz Reference
Users can correct the internal 10MHz reference signal using the 10MHz adjustment function
combined with a frequency counter.
Connect the
[10MHz OUT]
connector of the SSG to a frequency counter.
1. Restore factory settings
a) Press
UTILITY
or the touch screen to set
UTILITY
->
Settings
->
Preset Type
->
Default
.
b) Press
PRESET
, or press
UTILITY
->
Preset
.
2. Set the code of reference OCXO
a) Press
UTILITY
or the touch screen to set
UTILITY
->
Settings
->
10M Adjustment
->
,
enter the “Ref Osc setting”.
b) Press
10M Adjustment
, the switch will toggle to the open state, at which point the
OCXO will use the value in the
Ref Osc Code
as the code of the OCXO.
If the
reference calibration switch is off, the OCXO will use the default code.
c) Press the
Ref Osc Code
, a small keyboard will pop up, enter the code.
3.
Correcting 10MHz Reference
a)
According to the frequency measured by the frequency meter, if it is lower than 10MHz,
increase the
Ref Osc Code
, otherwise decrease the
Ref Osc Code
. Repeatedly
configure the
Ref Osc Code
until the 10MHz reference signal output by the OCXO meets
the accuracy requirements.