106
User Manual
4.1.2
Freq Offset
Set the frequency offset value to illustrate the frequency conversion between the measured device
and the input of the spectrum analyzer.
⚫
This parameter does not affect any hardware settings of the spectrum analyzer, but only
changes the display values of center frequency, start frequency and stop frequency.
⚫
To eliminate the frequency offset value, the frequency offset value can be set to 0 Hz.
Command Format
[:SENSe]:FREQuency:OFFSet
[:SENSe]:FREQuency:OFFSet?
Instruction
Sets the frequency offset of the spectrum analyzer.
Gets the frequency offset.
Parameter Type
Float, unit: Hz, kHz, MHz, GHz
Parameter Range
-
100 GHz ~ 100 GHz
Return
Float, unit: Hz
Example
:FREQuency:OFFSet 1 GHz
4.1.3
Freq Step
Setting the value of Freq Step will change the direction key step of center frequency, start
frequency, stop frequency and frequency offset.
⚫
At a fixed step change the value of the center frequency can reach the purpose of switching
measurement channels rapidly and continuously.
⚫
There are two kinds of frequency step modes
:
Auto
and
Manual
. In Auto mode, the Freq step
is 1/10 of the span in Non-zero span or equals the RBW while in Zero Span. In Manual mode,
you can set the step using the numeric keys.
Command Format
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Instruction
Specifies the center frequency step size.
Gets the center frequency step.
Parameter Type
Float, unit: Hz, kHz, MHz, GHz
Parameter Range
1 Hz ~ 40 MHz
Return
Float, unit: Hz
Example
:FREQuency:CENTer:STEP 2 MHz
Содержание SSA5083A
Страница 1: ...User Manual EN_01A SSA5085A SSA5083A Spectrum Analyzer...
Страница 2: ......
Страница 4: ......
Страница 29: ...User Manual XXV 10 2 TROUBLESHOOTING 225...
Страница 30: ......
Страница 97: ...User Manual 67...
Страница 253: ...User Manual 223 Power SYSTem POWer OFF SYSTem RESTart Buzzer DISPlay WINDow BEEP STATe DISPlay WINDow BEEP VOLUme...
Страница 259: ......