SIGLENT
SPD1168X
Quick
Start
9
7. Measured output values
8. Timer state
Output Inspection
1. Check the output voltage
(
1
)
Turn on the power and make sure the channel current
setting is not zero when the instrument has no-load.
(
2
)
Presson/off button, the supply should be working in
constant voltage (CV) mode. You can check the voltage
range by adjustingthe voltage setpointfrom the minimum
(0V) to the maximum value (16V).
2. Check the output current
(
1
)
Turn on the power and make sure the voltage setting is
not zero.
(
2
)
Connect the output terminals(short) with aninsulated wire
that can handle 10 A or more (18 AWG single core, for
example).
(
3
)
Activate the output by pressingthe on/off button. The low
impedance (shorted) output will cause the instrument to
enter current control (CC) mode. You can check the
current range by adjusting the current setpointfrom the
minimum (0A) to the maximum value (8A).
Fuse Replacement
The specifications of the fuse are relative to the actual input line
voltage, shown in the table below. You also can refer to the rear
panel “input power requirement”.
Input
voltage
Fuse
specification