SDS1000X User Manual
159
To Set Output Load
11. Press the
Wave Gen
button on the front panel to enter the WAVEFORM function
menu.
12. Press the
Setting
softkey to enter the SETTING function menu.
13. Press the
Output
Load
softkey to select
High-Z
or
50Ω
.
Figure 91 SETTING Function Menu
The output load of the Wave Gen must match with the oscilloscope, or the amplitude of
the signal will be incorrect.
Recover the Default Values of Wave Gen.
1. Press the
Wave Gen
button on the front panel to enter the WAVEFORM function
menu.
2. Press the
Setting
softkey to enter the SETTING function menu.
3. Press the
Default
softkey to recover the default values of Wave Gen (Sine, 1.000 KHz,
4.000Vpp, 0.0mVdc).