CP6000/CP4000
22 Current Probe Instructions
Getting Started
The CP4000 series current probe enables a general purpose oscilloscope to display
AC and DC current signals up to 100 amps Peak (70A RMS). The current probe can
also make AC and DC measurements with a multimeter by using the recommended
accessory MT-246(BNC-to- banana) plug adapter.
CP4000 series controls and indicators
Control/Indicator
Description
Current flow symbol.
The arrow shows the
probe’s polarity
convention for measuring current flowing from positive to negative.
Zero adjustment.
Rotate to adjust the probe output to zero when
there is no current present. It may also be used to offset a DC signal
component. Zeroing is not needed for AC measurements unless your
instrument cannot isolate a DC component( if present ).
OFF/Range switch.
Slide the switch from OFF to either the 10mV/A
or 100mV/A range. When either range is selected, the probe is turned
on, and the green battery indicator lights.
Battery indicator.
The green battery indicator lights when the probe
is turned on. For more information, see Battery Notes and Battery
Installation on Page 11.
Overload indicator.
The red overload indicator lights if the
measured signal is greater than the selected range capacity. Switch
the probe to 10 mV/A if possible, or remove the probe from the circuit.