SDS5000XSeries Digital Oscilloscope Service Manual
35/ 90
repeat step 4 to step6.
8.
Turn the 9500B output off.
9.
Disconnect the connections to all the channels.
Table 3
DC Gain Accuracy
Volt/Div
DC voltage output levels
DC Gain Accuracy error
10 V/div
30 V,15V ,0,-15 V,-30 V
±
1.5%
5 V/div
15 V, 7.5 V,0,-7.5 V,-15 V
±
1.5%
2 V/div
6 V,3 V,0,-3 V,-6 V
±
1.5%
1 V/div
3 V,1.5 V,0,-1.5 V,-3 V
±
1.5%
500 mV/div 1.5 V, 0.75 V,0,-0.75 V,-1.5 V
±
1.5%
200 mV/div 600 mV, 300 mV,0,-300 mV,-600 mV
±
1.5%
100 mV/div 300 mV,150 mV,0,-150 mV,-300 mV
±
1.5%
50 mV/div
150 mV,75 mV,0,-75 mV,-150 mV
±
1.5%
20 mV/div
60 mV,30 mV,0,-30 mV, -60 mV
±
1.5%
10 mV/div
30 mV,150 mV,0,-150 mV,-30 mV
±
1.5%
5 mV/div
15 mV,7.5 mV,0,-7.5 mV-15 mV
±
3%
2 mV/div
6 mV, 3 mV,0,-3 mV, -6 mV
±
3%
1 mV/div
3 mV, 1 mV,0,-1 mV, -3 mV
±
3%