Configuration and operation
4.1
Mounting
The V1x Geophone shall be mounted according to:
V10 shall be mounted either on the wall or on the floor, and
have the label parallel with the wall (readable from the side)
V11 shall be mounted either on the wall or on the floor, and
have the label parallel with the floor (readable from above)
V12 shall be mounted either on the wall or on the floor, and
have the label parallel with the floor (readable from above)
V12R shall be mounted in the ceiling and have the label
parallel with the ceiling (readable from below)
Note! For correct measuring results, the mounting angle of the V1x
Geophone must not exceed 5º relative to the reference plane.
Note! To facilitate proper communication in electrically harsh
environments, the metallic enclosure must not be in contact with any
kind of conductive surfaces (the sensor feet are isolated). All cable
junctions touching any metal or the ground shall be protected by
vulcanized tape.
Example of V10 and V12 Geophone mounting.