Quick Start
1.
Connect your S50 Sound Level Meter(s) with an INFRA
Datalogger using any of the available INFRA Sensor
Cables.
2.
Start the Datalogger and wait for the initialization process
to be finished.
3.
Under the menu
Nodes
you’ll find your S50 unit(s) as
SND S50-
<serial.no>
,
where SND is the customer string and
<serial.no>
is the
serial number.
4.
Select SND S50-
<serial.no>
in order to choose Standard
(S01, S02, etc.). Then, for the chosen standard, select
Interval time, Record time and Upper threshold level (value
and on/off). It is also possible to change the customer string.
Note! When doing a test or reference measurement, be sure
that the Upper threshold is set to OFF. Otherwise
unnecessary transients will be recorded.
5.
Activate REG ON, and the measurement will be initiated.
6.
Note!
First registration of interval data is delayed to
synchronize to the local time. The delay may be up to one
interval period. First presentation of data will then be
delayed up to two interval periods. Transient trig and
readout of Live-values are available directly after activation
of REG ON.