14
Chapter 4: Installing and Removing Flow Cells
The CalTrak accepts interchangeable CalTrak SL 800 specific flow cells for different flow ranges. When a cell is
installed,
the home screen displays the model number and flow range of the installed cell. If no cell is installed,
this information will be blank. The base includes a blue friction fit adapter ring that should be removed when using
the -75 cell
due to its larger diameter. Replace this adapter ring when using the other cells.
Installing Flow Cells
1.
Ensure blue adapter ring is installed for using all cells except the -75. Remove the ring if installing the -75 cell.
2.
Position the selected flow cell into the base opening, its top label facing you.
3.
Carefully align the cell with the guide pins; when the guide pins are engaged, press down.
4.
When powered on, the CalTrak SL 800 senses which cell is installed and displays the appropriate units for that
cell
.
Removing Flow Cells
Grasp the flow cell firmly, hold the base in place, and lift upwards.