Uroskop ACCESS
SPL5-330.840.01
Page 12 of 12
Siemens AG
Endoscopy Option
Rev. 01
03.03
CS SD 24
Medical Solutions
5 - 12
Error Correction
Check the brightness. A gray layer should be visible on dark (black) areas.
This may be a dark area in the image, or may be shaded at the optic input.
Image Sector Size
5
The circular image section may shift because of obsolete camera connectors (snap lock)
or connectors that have not been properly adjusted.
•
Replace the old endoscope.
Incorrect image sizes may result when using endoscopes from different manufacturers.
•
Use the endoscope provided by the camera manufacturer.