Operation
7.4 Operating modes
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
95
7.4.1
Warm-up phase
Figure 7-4
Warm-up phase, measuring mode, and input mode
Immediately following switching-on, the ULTRAMAT 23 tests
the display elements. During this test, all elements light up
simultaneously for approx. five seconds.
The adjacent display subsequently appears with the remaining
warm-up period which is counted down in seconds to 00:00
(minutes:seconds).
The analyzer initially carries out an AUTOCAL during the
warm-up phase. The flow of AUTOCAL gas (nitrogen or air) is
displayed in the bottom line, and the remaining time is shown
in the line above this. This calibration cannot be interrupted.
Following the calibration, the analyzer switches to purging with
sample gas. At the end of the purging phase, the analyzer
switches to measuring mode; however, full measuring accura-
cy is only reached after approx. 30 min if a further AUTOCAL
has been automatically carried out by the analyzer. The warm-
up phase is then finished.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...