Commissioning
6.5 System setup with several analyzers in parallel
ULTRAMAT 23
84
Manual, 3/2016, A5E37100388-003
The analyzer must have been in operation for at least 30 minutes before commencing with
measurements since a good stability of the analyzer unit is only guaranteed after this time
(99% value).
Make sure that the gas flow is between 1.2 and 2.0 l/min.
Any noise which may occur can be suppressed by adjusting various time constants (see
Parameters: Time constants (Page 129)).
The calibration should be repeated every six to twelve months depending on the ambient
conditions.
6.5
System setup with several analyzers in parallel
Example 1
Both analyzers with internal pump and solenoid valve switching between sample gas and
zero gas for AUTOCAL
The cyclic AUTOCAL of the master device triggers an AUTOCAL in parallel for the slave
device via its digital output SYNC and the digital input SYNC of the slave device.
The simultaneous connection between the digital output SYNC of the slave device and the
digital input SYNC of the master device guarantees that zero gas is always passed
simultaneously through both analyzers.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...