Safety instructions
2.1 Analyzers in biogas plants
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
17
2.1
Analyzers in biogas plants
DANGER
Danger of poisoning
This device is designed to measure hydrogen sulfide and dihydrogen sulfide, H
2
S)!
Hydrogen sulfide is highly toxic even in small concentrations! The odor threshold for
hydrogen sulfide is very low at 0.02 vpm (20 vpb), but higher concentrations result in
numbing of the olfactory receptors in the nose so that the odor is no longer perceived.
Persons exposed to this gas in concentrations up to 100 vpm for several hours exhibit
symptoms of poisoning such as fatigue, headaches, lack of appetite, lack of concentration,
irritation of the mucous membranes of eyes and respiratory tract, and throat irritations.
Inhalation of H
2
S concentrations of 500 vpm longer than 30 minutes can cause fatal
poisoning. Concentrations above 1 000 vpm cause death within a few minutes,
concentrations above 5 000 vpm cause death within a few seconds!
When using this device in plant where there may be high concentrations of H
2
S and you
therefore need to take following continual precautions to prevent the effects of poisoning:
•
Connect the gas outlet of the analyzer to a gas exhaust unit so that no gas can escape
into the environment!
•
Before you begin maintenance on the analyzer, make sure that the H
2
S concentration in
the analyzer is close to 0 vpm. Before beginning work, always flush the gas path of the
analyzer and the gas sampler with ambient air or nitrogen for a duration of about 10
minutes .
•
Check for leaks in the analyzer at regular intervals!
DANGER
Danger of explosion
This device is used in biogas plants, among other places. When it is used in biogas plants,
you should expect that the sample gas will contain methane, which forms explosive
mixtures with oxygen or air in certain concentrations. These conditions are possible with
certain operating states of the plant.
2.2
Analyzers in hazardous areas
WARNING
Unsuitable device for the hazardous area
Danger of explosion.
•
Only use equipment that is approved for use in the intended hazardous area and
labelled accordingly.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...