Functions
8.5 Automatically executed functions
ULTRAMAT 23
156
Manual, 3/2016, A5E37100388-003
8.5.2
Probe protection function
Definition of probe protection function
A value 1.1 times the specified range can be considered as the maximum continuous
concentration. Although a measurement above such a concentration is still correct, the probe
is damaged by longer exposure. The maximum continuous concentration remains constant
even when changing the measuring range.
The protection function is also implemented with the 50 vpm probe for compatibility reasons
even though its maximum continuous concentration is 12.5 vpm. The protection function is
executed above this value.
Execution of this function is the same for all probes. The function test is set during execution
of the protection function in order to signal that the displayed value is incorrect.
How the protection function works
The protection function is triggered if the continuous measured value of H
2
S is greater than
the maximum continuous concentration (110 % of full-scale value) in measuring mode for a
period of 3 seconds.
The following occurs when the protection function starts:
●
The H
2
S measured value display is set to "*****"
●
An "H" (H
2
S protection function running) is displayed in the measurement screen at the
right edge where the test letter "!" appears (fault no longer pending has been logged).
●
The zero gas valve is opened
●
The "Function control" status is set.
As long as the protection function is active, the zero gas valve remains open during the zero
gas purge time. The device then switches to purging of the sample gas path with sample
gas. This process is repeated if the maximum continuous concentration is exceeded within
the sample gas purge time.
This process is repeated up to 6 times. If after the 6th repeat the sample gas concentration is
still too high, the zero gas valve remains permanently open and the fault 28 "H
2
S probe
protection" is entered in the logbook.
If the maximum continuous concentration is not exceeded again, the protection state is
terminated and the H
2
S measured value is displayed again. Furthermore, the function control
and the test letter "H" are deleted.
Return to measuring mode
An active protection function can be interrupted as follows:
●
Automatic: Prior to completion of the 6th repeat, the measured value in the sample gas
purge time remains permanently below the maximum continuous concentration.
●
Set the protection function to 'OFF' by changing the parameter in the limit display or via
ELAN
●
Start a different state such as calibrate, AUTOCAL, etc.
●
Acknowledgment of the fault "H
2
S probe protection" in the logbook
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...