Functions
8.4 Configuration
ULTRAMAT 23
152
Manual, 3/2016, A5E37100388-003
8.4.3.2
Configuration: Device test: Display/keys/flow
You can select the following three tests in this menu.
•
Test of display
In this test, all characters in the character set of this analyzer
are output in succession at every position of the display. The
display remains empty if characters are output which cannot
be displayed. This is a cyclic test, i.e. once the complete
character set has been processed, it starts from the
beginning again. The test is repeated continuously until
terminated by pressing the <ESC> key.
A corresponding message will inform you of this before the
test is started.
•
Test of keys
This test takes 30 s, and the time remaining up to the end of
the test is output on the display. During this time you can
press all input keys in succession. The analyzer normally
recognizes that a key has been pressed and indicates this.
This test cannot be prematurely cancelled.
•
Test of flow switch
This indicates whether the sample gas flow is correct or not.
Depending on the type of gas supply, it may be necessary to
switch on the pump using the <PUMP> key.
8.4.3.3
Configuration: Device test: Inputs/outputs
In this menu you can call the tests of the electric inputs and
outputs of the analyzer. You require the following equipment to
carry out these tests:
•
Ammeter
•
Ohmmeter
•
Power supply (24 V DC)
•
Test plugs
Configuration: Device test: Inputs/outputs: Test analog outputs
The analyzer has four analog outputs and an output current
range of 0/2/4 ... 20 mA. You can test these by setting any val-
ue of the output current between 0 and 20 mA in this menu dis-
play.
To test these outputs, connect an ammeter to the correspond-
ing analog outputs on plug X80 and measure the output current.
The pin assignments of plug X80 are described in section Pin
assignments (Page 54).
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...