Operation
7.4 Operating modes
ULTRAMAT 23
102
Manual, 3/2016, A5E37100388-003
The <CAL> key cannot be used during the warm-up phase.
If the flow is too low during a zero calibration triggered by pressing the key, the analyzer
remains in this status until either the flow is sufficient or the zero calibration is aborted by
pressing the <ESC> key.
In addition to the <CAL> key, an AUTOCAL can also be triggered via the binary input. The
binary input has priority over the key.
7.4.3.5
The PUMP key
If the analyzer is equipped with an internal sample gas pump, this can be switched on and
off using the <PUMP> key. If the pump is switched off while the analyzer is in input mode, it
is also switched on again by pressing the <MEAS> key if parameterized accordingly (see
Configuration: Inputs/outputs/pump: Pump at CAL/MEAS (Page 142)).
In addition to use of the <PUMP> key, the pump can also be switched on and off via the
binary input. The binary input has priority over the key.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...