Removing and Installing the Black and White Printer
3.0 Removing and Installing the Black and White
Printer
Use the instructions contained in this section to remove and install the black
and white printer (P91W) in the G50 and the G60 S ultrasound system.
Replacement of the black and white printer requires approximately 20 minutes.
3.1 Tools Required
Use a number 2 Phillips screwdriver to perform this procedure.
3.2 Removing the Black and White Printer
Perform the following steps to remove the black and white printer (P91W) from
the G50 or the G60 S system.
1.
Turn off the system and unplug the power cable from the outlet.
2.
Remove the two M3 X 8 binding head screws and external lock washers
from bracket F located under the front of the printer, shown in Figure 4-10,
and disconnect all of the cables from the rear of the P91W as you extract
the P91W from the OEM bay.
!
Warning:
The replacement procedure described in this section involves the removal and re-
installation of threaded fasteners. The installation of over-long fasteners is some locations can
penetrate and damage the underlying equipment or short underlying circuits energized at 90
Vac to 264 Vac. The potential damage can affect equipment performance; the potential shorting
of the electrical circuits can produce shock hazards that are capable of causing serious injury
or death. Verify that threaded fasteners of the correct length are installed in the equipment.
When fasteners are removed, store them with the related components for re-installation. Refer
to this procedure to verify that only fasteners of the correct length are re-installed. Re-install
threaded fasteners slowly, and verify that they do not contact underlying equipment.
!
Warning:
The G50 and the G60 S contain hazardous voltages that can range from 90 Vac to
264 Vac. These voltages can cause serious injury or death. Only authorized service personnel
are permitted to remove protective covers and access the energized components of this sys-
tem.
Содержание SONOLINE G50
Страница 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Страница 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Страница 182: ...Page 42 of 42 7482255 Rev 01 Siemens October 2002 Service Screens and Menus Figure 8 32 Color Scale Pattern ...
Страница 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Страница 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Страница 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Страница 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Страница 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Страница 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Страница 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Страница 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...